Your shopping cart is empty!
Ground Continuity
|
|
---|---|
Method
|
2 wire ZAP technique
|
Test Current
|
>± 200mADC into 2Ω
|
Max Test Voltage
|
4-24Vrms o/c
|
Measuring Range (low range) Resolution
|
0.001 - 0.999Ω
0.001Ω |
Measuring Range (mid range) Resolution
|
1.00 - 9.99Ω
0.01Ω |
Measuring Range (high range) Resolution
|
10.0 - 19.9Ω
0.1Ω |
Accuracy
|
±1% of value, ± 5mΩ
|
Circuit Protection
|
Test inhibited if ≥ 30VAC or DC at 4mm inputs
|
Voltage Measurement
|
|
---|---|
Application
|
H-N, H-G, N-G and touch voltage (IEC 61010)
|
Range
|
0.0V - 300VAC
|
Resolution
|
0.1Hz
|
Accuracy
|
± 2% ± 5 digits (between 10V - 270VAC)
|
Mains Frequency
|
45.0 - 66.0Hz
|
Power Source
|
|
---|---|
Maximum Current Rating
|
20A @ 120V / 16A @ 230V
|
Duty cycle (@70°F ambient)
|
16A to 20A, 3 min. on/ 10 min. off
10A to 15A, 3 min. on/ 5 min. off 0A to 10A, continuous |
Applied Parts Connection Box
|
|
---|---|
Size (L x W x D)
|
6.1 x 2.6 x 1.1” (not including flying lead)
|
Weight
|
0.33lbs
|
Rating
|
Category I
|
Environmental Protection
|
IP 40
|
Impact rating
|
5J
|
Meets the safety standard
|
BS EN 61010-1:2010 Pollution degree 2
|
Voltage rating (maximum)
|
230Vac
|
Powered Leakage Measurements
|
|
---|---|
NFPA-99
|
Ground leakage
Chassis leakage (Touch Current) Patient lead leakage |
Measuring Range
|
0.1 - 9999μA
|
Measurement/Display Resolution
|
0.1μA
|
Accuracy
|
± 2%, ±5μA
|
Mains Reversal
|
Soft key
|
Single Fault Conditions Open Neutral
|
Via soft key
|
Open Ground
|
Via soft key
|
EUT Load Current Measurement |
---|