WARRANTY: Three year Manufacturer warranty
LEAD TIME: 10-14 Day.
FUNCTIONS INCLUDE: NIBP, IBP, Arrhythmia, ECG, Pacer, Resp, Peak Detect, Alarm Test, Leak Detect
DESCRIPTION:
The SimCube SC-5 combines the functions of both a manometer (pressure gauge) and an NIBP simulator. The SC-5 NIBP function has five modes: Manometer (+/- 400mmHg), Adult (120/80 mmHg), Neonatal (70/40 mmHg), Hypertensive (190/120 mmHg), and Hypotensive (80/40 mmHg). A single button toggles through each of the modes. A mini-DIN connector provides access to 3 dynamic IBP values (120/80, 70/40, 190/120) and 7 static IBP values (0, 25, 50, 100, 150, 200, 250). Ten snaps provide 3-, 5-, or 12-lead electrocardiogram (ECG) simulation at 3 heartrates (70, 95, Asystole) plus a Heartrate sequence, as well as Arrhythmia, Pacer, and 3 Resp rates with a Resp Sequence. An Alarm Test mode toggles the ECG Heart Rate and Respiration values between 47/00 bpm and 160/80 bpm every 30s. Peak Detect (Overpressure) measurement and Leak Rate measurement functions are provided. The SC-5 is compatible with the Battery Boost Option.
TECHNICAL SPECIFICATIONS:
Physical Dimensions: Size 3x 3 x 3.5 (7.6cm X 7.6cm X 8.9cm)
Weight: 2.5 Lbs
Power External A/C Adaptor: (Output: 6VDC / 2amps, 2.1mm, center positive connector) or 4 AA Batteries (with Battery Boost Option)
NIBP Connection: Quick Disconnect, Female
ECG/Resp Connection : 10 ECG snaps
IBP Connection: Mini-DIN
Manometer Range: – 400 to + 400 mmHg
Precision: 0.1 mmHg
Accuracy: +/- 1 mmHg
User Interface: Single Button Operation
OPERATING MODES:
Adult NIBP, Neo NIBP, Hypertensive NIBP, Hypotensive NIBP, Manometer, Peak Detect, HR Seq., Alarm Test, ECG Pace ON,
Arrhythmia Sequence, Heart Rate Sequence, Invasive BP, Zero Invasive BP, 100, 200 Invasive BP Sequence
NIBP Adult Simulation:
Simulated Pressure: 120/80 (97) mmHg
Simulated Heart Rate 70 bpm
Simulated Pulse Volume 1 ml
NIBP Neonatal Simulation
Simulated Pressure: 70/40 (52) mmHg
Simulated Heart Rate: 95 bpm
Simulated Pulse Volume: 0.25 ml
NIBP Hypertensive Simulation
Simulated Pressure: 190/120 (146) mmHg
Simulated Heart Rate: 70 bpm
Simulated Pulse Volume: 1 ml
NIBP Hypotensive Simulation
Simulated Pressure: 80/40 (58) mmHg
Simulated Heart Rate: 70 bpm
Simulated Pulse Volume: 1 ml
ECG Simulation
Isolated: Yes
Synchronized with NIBP: Yes
R Wave Size: 1mV (lead I) +/- 5%
R Wave Width: 35 ms
Wave Shape: QRS wave
Connection : 10 Snaps
Simulation Rates: 70, 95, Asystole, Arrhythmia, Pacer, HR seq.
HR Sequence: 30 seconds each of:30, 60, 90, 120, 45, 160, and 220 bpm
Pacer Simulation
Isolated:Yes
Synchronized with NIBP: Yes
Pacer Size: 3 mV
Pacer Width : 1.2 ms
Respiration Simulation
Isolated : Yes
Synchronized with NIBP: Yes
Wave Shape: Square Wave
Size: 4 Ohm
Simulation rates: 20, 40, Apnea, Sequence = 00, 30, 45, 60, 22, 30, 80, 110.
Arrhythmia Simulation
Cardiac failure sequence: ~ 90 seconds of normal beats interspersed with PVCs and Runs, followed by ~30 seconds of VTAC, followed by ~30 seconds of VFIB, concluding with ~30 seconds of asystole.
Peak Detect
Precision 1.0mmHg
Invasive Blood Pressure Simulation
Isolated: Yes
Synchronized with NIBP: Yes
Excitation Voltage DC range = 3.3 to 5.8 AC range = 3.3 to 12p-p
Pressure range: 0-250 mmHg
Simulated Pressure accuracy: +/- 1 mmHg
Wiring + Excit = pin 1, - Excit = pin 4, + Sig = pin 3, -Sig = pin 6
Simulation rates Dynamic = 120/80, 70/40, 190/120. Static = 0, 100, 200. Step = 0,25,50,100,150,200, 250
Environmental
Voltage Range: 100-240 VAC, 50-60 Hz