"NEW" Netech MiniSim 1000 Multiparameter Simulator
|
|
|
"NEW UNIT" Comes with manufacturer one-year warranty
MiniSim Model 1000 Multi-Parameter Patient Simulator
The MiniSim 1000 is a powerful, comprehensive patient simulator with a number of unique features. It provides a full 12 lead ECG signal with more than 40 arrhythmias, dual blood pressure, respiration, and temperature. It also delivers square, sine, and a triangle performance waveforms.
MiniSim 1000 arrhythmia selections include pacer waveforms, ST segment changes, and both atrial and ventricular waveforms. The dual blood pressure and respiration waveforms are synchronized to the ECG and track with all arrhythmia selections. Multiple static and dynamic pressure readings are selectable as per apnea events
Operation of the hand - held MiniSim 1000 is simple via menu driven tactile keys utilizing a large LCD Display. The instrument may be either AC or Battery operated.
Testing Features:
12 Lead ECG Over 40 Arrhythmias ECG Performance Waveforms Dual Blood Pressure Respiration Temperature
Operating Features:
Easy to Use Small, Compact Size Menu Driven Operation Large LCD Display Tactile Keypad Battery or AC Operation Specifications: ECG Normal Sinus: Rate: 30, 60, 70, 80, 90, 100,120, 150, 180, 210, 240, 270, 300, 350 BPM Amplitude: 0.15 to 5mV High Level Output: 0.5 Volt/mV of Low Level (Lead II)
Performance Waveforms: Sine, Triangle, Square Rate: 0.1 to 100 Hz 0.1 to 1 Hz in 0.1 Hz increments 1 to 10 Hz in 1 Hz increments 10 to 100 Hz in 10 Hz increments
Tall T Wave
Test Waveform@ 70 BPM Atrial Pacer Ventricular Pacer* Paced Rhythms - Variable from 0.1 to 2mS R Wave detection - Variable form 10 to 120mS.
Pulse: 2mS Pulse of 1mV amp @ 4 sec intervals
Pacers ST Segment Analysis: ST Elevation and Depression ST Depression Positive and Negative Slope Myocardial Infarction
Arrhythmias: Atrial Ventricular Blocks
Respiration: Normal Physiological Simulation: Base Line Impedance: 250, 500, 750, 100 Ohms Delta Impedance: 0.1, 0.5, 1.0, 1.5 Ohms Rates: 15, 30, 60, 120 BPM Apnea: Off, Continuous, 12 Seconds and 32 Seconds Apnea-Bradycardia Accuracy: Delta Impedance 5% of selection Rate: 1% of selection
Blood Pressure: Impedance: 350 Ohms Excitation: 2 to 16 Volts Sensitivity: 5uV/V/mmHg Static: 0, 5, 10, 20, 25, 30, 40, 50, 100, 150, 200, 300 mmHg. Dynamic: 50/10, 60/20, 70/30, 80/40, 100/60, 120/80 mmHg. Accuracy: +/-2% FS. +/-1 mmHg Synchronized with all normal sinus rates Physiologically track all arrhythmia selections.
Temperature: Compatibility: YSI 400/700 25/37 Degree Centigrade Accuracy: +/-1% of reading
General: Display: 16 Character 2 Line LCD Key Pad: 8 tactile keys Ð F1, F2, F3, F4, >, <, ON/OFF, Menu Output Connectors ECG: 10 Banana Jacks High Level: 3.5mm Phono Jack Pressure: 5 contact circular Din Temperature: 3.5mm Stereo Jack Physical Size: 5 1/2 x 31/2 x 1 1/2 inches Case: High Impact Plastic Weight: 10 Oz.
Power: 9 Volt Alkaline/AC adapter (Optional). Battery Life 40 Hours Continuous. Display Indicates Low Battery.
|
|

|
|