PREOWNED Fluke 180 Electrical Safety Analyzer

PREOWNED Fluke 180 Electrical Safety Analyzer
PREOWNED Fluke 180 Electrical Safety Analyzer PREOWNED Fluke 180 Electrical Safety Analyzer
$995.00
Ex Tax: $995.00
Product Code: 180
Availability: In Stock
WARRANTY: One Year
SALES LEAD TIME: 14 DAYS
ACCESSORIES: Chassis Test Lead, Soft Carrying Case, Calibration Certificate
 
OVERVIEW:
The handheld 180 Electrical Safety Analyzer is a lightweight, portable device for testing electrical systems, medical devices, and physiological instrumentation. The device includes the AAMI test loadand has 5 jacks for patient applied parts testing. Small enough to fit in a briefcase, the analyzer works well as a bench-top instrument in the laboratory or a portable testing device in the mobile engineer’s toolbox. With its uncomplicated design, the 180 is simple to use. A single master function switch, directly labeled with the test to be performed, leads the user through a complete measurement procedure. The analyzer uses simple, yet sophisticated, electronics for true RMS measurement of current and voltage. The 180 also performs dual-lead leakage and resistance tests.
 
KEY FEATURES:
  • Handheld | Self-switching 120 V/240 V operation | 15 A and 20 A capabilities | FIVE patient applied parts jacks | Dual-lead testing
 
SPECIFICATIONS:
Dimensions
5.25 in L x 7.25 in W x 2.1 in H (13.3 cm x 18.4 cm x 5.4 cm)
Weight
Max weight 2.25 lb (1 kg)
Display
LCD 3.5 digit
Operating Mains Voltage Range
90 V ac to 240 V ac
Current Capacity for DUT
  • Line 90V to 140V: 20A for 5 min, 15A for 30 min
  • Line over 140V: 10A
Line-Voltage Measurement
Range:
90 to 240 V
Accuracy:
±3 % of reading 1 LSD
Load-Current Measurement
Range:
1.00 to 19.99 A
Accuracy:
±5 % of reading 1 LSD
Leakage-Current Measurement
Range:
0 to 1999 µA
Accuracy:
  • DC and 25 Hz to 1 kHz 1 % of reading 3 LSD
  • 1 to 100 kHz 2.5 % of reading 3 LSD
  • 100 kHz to 1 MHz 5.0% of reading 3 LSD
Measurement type:
True RMS
Input impedance per AAMI ESI 1993
Isolation Test
Voltage:
110 % of mains, 5 % of reading
Current limit:
1 mA
Resistance Measurement
Range:
0 to 2.00 Ω, 2.01 to 19.99 Ω
Accuracy:
±1 % of reading 1 LSD, ±3 % of reading 1 LSD
Resolution:
0.01 Ω
Current source:
10 mA DC