IMTMedical Citrex H4 Portable Gas Flow Test Kit

IMTMedical Citrex H4 Portable Gas Flow Test Kit
IMTMedical Citrex H4 Portable Gas Flow Test Kit IMTMedical Citrex H4 Portable Gas Flow Test Kit IMTMedical Citrex H4 Portable Gas Flow Test Kit IMTMedical Citrex H4 Portable Gas Flow Test Kit IMTMedical Citrex H4 Portable Gas Flow Test Kit IMTMedical Citrex H4 Portable Gas Flow Test Kit IMTMedical Citrex H4 Portable Gas Flow Test Kit
$5,290.00
Ex Tax: $5,290.00
Brand: IMT Medical
Product Code: CITREX H4
Availability: In Stock
SALES LEAD TIME: 14-21 DAYS
WARRANTY: One Year
ACCESSORIES: CITREX H4 • EasyLung • CITREX Webserver configurator and monitoring • FlowLab activation code • Oxygen sensor incl. activation code • Adapter set • Laminar flow tube • Car adapter • Micro SD memory card • Protection filter RT019 • USB cable • Ethernet cable • Power supply • Carrying bag • Battery pack • Quick start manual
 
GENERAL:
The CITREX H4 test set is a certified, high precision measuring system with an extensive set of accessories. The CITREX H4 can be used to validate and calibrate a wide range of ventilators and gas delivery systems. It’s intuitive user interface and high resolution display makes set up and measurement a simple process.
 
CITREX H4 The device measures bi-directional flow, pressure, temperature and oxygen concentration. CITREX H4 is an appealing device thanks to its compact design, low weight and robustness. CITREX Webserver Monitoring Connect CITREX H4 to your IT network. Remote monitoring via your LAN network or the internet is child's play. The “Webserver Monitoring” software tool included in the scope of delivery can be used in any browser to evaluate and record your measured data in detail. FlowLab The test set also includes the advanced FlowLab software for detailed analysis and recording of all CITREX parameters. High resolution real-time curves, trending capabilities and advanced Excel reporting functions make it indispensable to experts.
 
ACCESSORY KIT CONTAINS:
CITREX H4 test set contains: • CITREX H4 • EasyLung • CITREX Webserver configurator and monitoring • FlowLab activation code • Oxygen sensor incl. activation code • Adapter set • Laminar flow tube • Car adapter • Micro SD memory card • Protection filter RT019 • USB cable • Ethernet cable • Power supply • Carrying bag • Battery pack • Quick start manual Differential pressure In addition to 17 gas standards and 8 gas types, the CITREX H4 has also a differential pressure sensor with a range of ±200 mbar. The measured values of this sensor can be used for internal calculation of ventilation parameter and stored on an external memory media.
 
WHAT THE CITREX H4 CAN BE USED TO TEST:
CITREX H4 test set for inspecting and certifying: • CPAP/ bi-level ventilators • ICU ventilators • High-frequency ventilators • Blood pressure monitors • Oxygen concentrators • Spirometers • CO2 insufflators • Medical gas supplies
 
SPECIFICATIONS:
Flow and Pressure Measurements Range Accuracy
Flow ± 300 sL/min*** Air: ± 1.9%* or ± 0.1 sL/min**
Temperature compensated yes
Pressure compensated yes
Pressure High 0...10 bar ± 1%* or ± 10 mbar**
Differential ± 200 mbar ± 0.75%* or ± 0.1 mbar**
Flow channel - 50...150 mbar ± 0.75%* or ± 0.1 mbar**
Units Flow L/min, L/s, cfm, mL/min, mL/s Pressure bar, mbar, cmH2O, inH2O, Torr, inHg, hPa, kPa, mmHg,
PSI Other Measurements Range Accuracy Oxygen, pressure compensated 0...100% ± 1% O2**
Gas temperature 0...50°C ± 1.75%* or ± 0.5°C**
Gas types Air, Air/O2, N2O/O2, Heliox (21% O2), N2, CO2
Gas standards ATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPS-A, BTPD, BTPD-A, 0/1013, 20/981, 15/1013, 25/991, 20/1013, NTPD, NTPS
Ventilation Parameters Range Accuracy Breath rate 1...1000 bpm ±1 bpm or ± 2.5%**
Time Ti, Te 0.05...60 s ± 0.02 s
Ratio I:E 1:300...300:1 ± 2.5%*
Ti/Ttotal 0...100% ± 5%*
Volume VTi, Vte ± 10 L ± 2%* or ± 0.20 mL (> 6 sL/min)**
Minute volume Vi, Ve 0...300 sL/min ± 2.5%*
Peak flow Insp./Exp. ± 300 sL/min Air: ± 1.9%* or ± 0.1 sL/min**
Pressure Ppeak, Pmean, PEEP, Pplateau 0...150 mbar ± 0.75%* or ± 0.1 mbar**
Compliance Cstat 0...1000 mL/mbar ± 3%* or ± 1mL/mbar**
Volume trigger Adult, Pediatric, HFO flow or pressure at preset and at adjustable levels
General Information
Color display yes
Realtime curves flow, pressure, volume, temperature, oxygen, ventilation parameters
Interface RS-232, USB, Ethernet, CAN, Analog Out (TTL), TSI4000 Protocol
Power 100...240 VAC, 50...60 Hz
Battery 4 hours
Dimension (wxdxh) 11.4 x 6 x 7 cm
Weight 0.4 kg
Calibration annually
Memory card yes
Approvals CE, CAN/CSA-C22.2 No. 61010-1-12, EN 61326-1:2006/ICE 61326-2:2005 (EMC)