Welcome visitor you can login or create an account.

Currency: US Dollar

BIOMEDEQUIP, INC.

0 item(s) - $0.00

Your shopping cart is empty!

Fluke ESA 175 Electrical Safety Analyzer

Fluke ESA 175 Electrical Safety Analyzer
Fluke ESA 175 Electrical Safety Analyzer Fluke ESA 175 Electrical Safety Analyzer
$895.00
Ex Tax: $895.00
Brand: Fluke Biomedical
Product Code: ESA 175
Availability: In Stock
"THIS IS A PREOWNED UNIT"
SALES LEAD TIME: 14 DAYS (unit will be sent in for fresh calibration when order is placed)
WARRANTY: One year
ACCESSORIES: Soft Carrying Case, Operator Manaul (CD), Chassis Test Cable, Fresh Calibration Certificate
 
INTRODUCTION:
The 175 Electrical Safety Analyzer is ideal for performing quick electrical safety checks on electrical systems, medical devices, and physiological instrumentation. Small enough to fit in a briefcase, the analyzer works well as a bench-top instrument in the laboratory or a portable testing device in the mobile engineer’s toolbox. With its uncomplicated design, the 175 is simple to use. A single master function switch, directly labeled with the test to be performed, leads the user through a complete measurement procedure. The unit features both IEC601-1 and AAMI test loads. Technicians simply flip a switch to perform leakage measurements to a particular standard. The analyzer uses simple, yet sophisticated, electronics for true RMS measurement of current and voltage. The 175 also performs dual-lead leakage tests. 
 
FEATURES:
Handheld
Self-Switching 120V/240V Operation
15A for 10 minutes
Dual-Lead Testing
 
SPECIFICATIONS:
Operating Mains Voltage Range:  90VAC to 240VAC
Current Capacity for DUT:  Line 90VAC to 140VAC: 15A for 10min; line over 140V: 10A
Line Voltage Measurement:
Range:  90VAC to 240VAC
Accuracy:  +/- 3% of reading +/- 1 LSD
Load Current Measurement:
Range:  1A to 19.99A
Accuracy: +/- 5% of reading +/- 1 LSD
Leakage Current Measurement:
Range:  0uA to 1999uA
Accuracy: DC and 25Hz to 1kHz: +/-1% of reading +/- 1uA; 1kHz to 100kHz: +/-2.5% of reading +/- 1uA; 100kHz to 1MHz: +/-5% of reading +/- 1uA
Measurement Type:  True-rms
Isolation Test:
Isolation Source Voltage: 110% of mains, +/- 5% of reading
Current Limit: 1mA @ 120VAC
Resistance Measurement:
Range: -.01ohms to 19.99ohms
Accuracy: +/- 1% of reading +/- 1 LSD
Resolution: 0.01ohms
Current Source: 10mA dc
Environmental Requirements:
Operating Temperature: 15degrees C to 40degrees C (59degrees F to 104degrees F)
Storage Temperature:  -20degrees C to 65degrees C (-4degrees F to 149degrees F)
Relative Humidity: 90% max
Mains Voltage Range: 90VAC to 240VAC
General Information:
Display: LCD 3.5 digit
DIMS: 13.3cm x 18.4cm x 5.4cm (5.25in x 7.25in x 2.1in)
Weight: 1kg (2.25lb)