SC-5 Kit NIBP Mini-Tester Complete Kit w/IBP, ECG, Pacer, Resp, Battery Boost

SC-5 Kit NIBP Mini-Tester Complete Kit w/IBP, ECG, Pacer, Resp, Battery Boost
Call for Price
Brand: Pronk Technologies
Product Code: Pronk Technologies SC-5 Mini NIBP Tester Kit
Availability: In Stock

WARRANTY: Four-year Manufacturer warranty
SALES LEAD TIME: 10-14 Day
ACCESSORIES INCLUDED: The SC-5KIT includes an SC-5 with all NIBP adapters (5), the Battery Boost Option, Cuff Jacket Duo, IBP extension cable (not shown), and one IBP Adapter Cable of your choice (not shown). Operator Manual, Calibration Certificate

FUNCTIONS INCLUDE: NIBP, IBP, Arrhythmia, ECG, Pacer, Resp, Peak Detect, Alarm Test, Leak Detect 

DESCRIPTION OF THE BATTERY BOOST OPTION:
The Battery Option (shown here as part of the SC-5KIT) includes both our Battery Boost module and a specially-modified carrying case. The Battery Boost module produces the 6V supply required by the SimCube from a set of 4 "AA" alkaline or NiMH batteries. The 6V output is maintained even as the battery voltage drops from 6V down to around 3V during discharge, providing for 150-250 accurate NIBP simulations from a single set of batteries. The case also provides storage for all NIBP adapters as well as the Bulb Adapter, spare batteries, and AC adapter. The Battery Boost module slides easily out of the case for quick battery replacement. The AC adapter is stored in a convenient outside zip pocket, and the Battery Boost module automatically switches to AC operation whenever the unit is plugged in. AC status and battery status are shown on the Battery Boost module LED display. Compatible with SC-3, SC-4 and SC-5.

DESCRIPTION OF THE SC-5:
The SimCube SC-5 combines the functions of both a manometer (pressure gauge) and an NIBP simulator. The SC-5 NIBP function has five modes: Manometer (+/- 400mmHg), Adult (120/80 mmHg), Neonatal (70/40 mmHg), Hypertensive (190/120 mmHg), and Hypotensive (80/40 mmHg). A single button toggles through each of the modes. A mini-DIN connector provides access to 3 dynamic IBP values (120/80, 70/40, 190/120) and 7 static IBP values (0, 25, 50, 100, 150, 200, 250). Ten snaps provide 3-, 5-, or 12-lead electrocardiogram (ECG) simulation at 3 heartrates (70, 95, Asystole) plus a Heartrate sequence, as well as Arrhythmia, Pacer, and 3 Resp rates with a Resp Sequence. An Alarm Test mode toggles the ECG Heart Rate and Respiration values between 47/00 bpm and 160/80 bpm every 30s. Peak Detect (Overpressure) measurement and Leak Rate measurement functions are provided. The SC-5 is compatible with the Battery Boost Option. 

TECHNICAL SPECIFICATIONS:
Physical Dimensions: Size 3”x 3” x 3.5” (7.6cm X 7.6cm? X 8.9cm) 
Weight: 2.5 Lbs 
Power External A/C Adaptor: (Output: 6VDC / 2amps, 2.1mm, center positive connector) or 4 AA Batteries (with Battery Boost Option) 
NIBP Connection: Quick Disconnect, Female 
ECG/Resp Connection: 10 ECG snaps 
IBP Connection: Mini-DIN 
Manometer Range: 400 to + 400 mmHg 
Precision: 0.1 mmHg 
Accuracy: +/- 1 mmHg 
User Interface: Single Button Operation 

OPERATING MODES:
Adult NIBP, Neo NIBP, Hypertensive NIBP, Hypotensive NIBP, Manometer, Peak Detect, HR Seq., Alarm Test, ECG Pace ON, 

Arrhythmia Sequence, Heart Rate Sequence, Invasive BP, Zero Invasive BP, 100, 200 Invasive BP Sequence 

NIBP Adult Simulation: 
Simulated Pressure: 120/80 (97) mmHg 
Simulated Heart Rate 70 bpm 
Simulated Pulse Volume 1 ml 

NIBP Neonatal Simulation 
Simulated Pressure: 70/40 (52) mmHg 
Simulated Heart Rate: 95 bpm 
Simulated Pulse Volume: 0.25 ml 

NIBP Hypertensive Simulation 
Simulated Pressure: 190/120 (146) mmHg 
Simulated Heart Rate: 70 bpm 
Simulated Pulse Volume: 1 ml 

NIBP Hypotensive Simulation 
Simulated Pressure: 80/40 (58) mmHg 
Simulated Heart Rate: 70 bpm 
Simulated Pulse Volume: 1 ml 

ECG Simulation 
Isolated: Yes 
Synchronized with NIBP: Yes 
R Wave Size: 1mV (lead I) +/- 5% 
R Wave Width: 35 ms 
Wave Shape: QRS wave 
Connection : 10 Snaps 
Simulation Rates: 70, 95, Asystole, Arrhythmia, Pacer, HR seq. 
HR Sequence: 30 seconds each of:30, 60, 90, 120, 45, 160, and 220 bpm 

Pacer Simulation 
Isolated:Yes 
Synchronized with NIBP: Yes 
Pacer Size: 3 mV 
Pacer Width : 1.2 ms 

Respiration Simulation 
Isolated : Yes 
Synchronized with NIBP: Yes 
Wave Shape: Square Wave 
Size: 4 Ohm 
Simulation rates: 20, 40, Apnea, Sequence = 00, 30, 45, 60, 22, 30, 80, 110. 

Arrhythmia Simulation 
Cardiac failure sequence: ~ 90 seconds of normal beats interspersed with PVCs and Runs, followed by ~30 seconds of VTAC, followed by ~30 seconds of VFIB, concluding with ~30 seconds of asystole. 

Peak Detect 
Precision 1.0mmHg 

Invasive Blood Pressure Simulation 
Isolated: Yes 
Synchronized with NIBP: Yes 
Excitation Voltage DC range = 3.3 to 5.8 AC range = 3.3 to 12p-p 
Pressure range: 0-250 mmHg 
Simulated Pressure accuracy: +/- 1 mmHg 
Wiring + Excit = pin 1, - Excit = pin 4, + Sig = pin 3, -Sig = pin 6 
Simulation rates Dynamic = 120/80, 70/40, 190/120. Static = 0, 100, 200. Step = 0,25,50,100,150,200, 250 

Environmental 
Voltage Range: 100-240 VAC, 50-60 Hz