Pronk Technologies SC-4 Mini NIBP Tester

Pronk Technologies SC-4 Mini NIBP Tester
Call for Price
Brand: Pronk Technologies
Product Code: Pronk Technologies SC-4 SimCube NIBP Tester
Availability: In Stock

SALES LEAD TIME: 14 Days
ACCESSORIES INCLUDED: AC Power Adapter, one tubing adapter, calibration certificate.

WARRANTY: Four year warranty on unit

SC-4 SimCube Oscillimetric NIBP Simulator w/ECG/Resp/Pacer, Automated Alarm Test Mode and Peak Detect.

TEST MODES
NIBP Manometer:   Static Manometer readout displays pressures from  0 - 480mmHg in 0.5 mmHg steps . This mode is used to test for calibration and leak problems. NIBP output is OFF.

NIBP Peak Detect: Readout will latch the highest pressure value sensed, in 0.5 mmHg steps. The readout value is reset when the SimCube senses a static 00mmHg for 10 seconds.

ECG HR/RESP Alarm Test: The ECG Heart Rate and Respiration values will toggle every 30 seconds. ECG HR/RESP values toggle between 47/00 bpm and 160/80 bpm. NIBP Manometer is automatically selected.

ECG HR/RESP Pacer ON: ECG output rate is 70 bpm with pacer artifact. RESP output rate is 35 bpm. This mode can be used to test pacer detection. NIBP Adult is automatically selected.

NIBP Adult (120/80): Continuous NIBP adult simulation is generated in response to detected air pressure with a pulse envelope that simulates a patient with blood pressure of 120/80 mmHg, heart rate of 70 bpm, and mean pressure of 100 mmHg, with simulated pulse volume of 1ml. Requires an adult cuff in-line with SimCube.

NIBP Neo (70/40): Continuous NIBP Neonatal simulation is generated in response to detected air pressure with a pulse envelope that simulates a patient with blood pressure of 70/40 mmHg, heart rate of 94 bpm, and mean pressure of 55 mmHg, with simulated pulse volume of 0.5ml. Requires an neo cuff in-line with SimCube.

NIBP High (190/120): Continuous NIBP Hypertensive simulation is generated in response to detected air pressure with a pulse envelope that simulates a patient with blood pressure of 190/120 mmHg, heart rate of 70 bpm, and mean pressure of 150 mmHg, with simulated pulse volume of 1ml. Requires an adult cuff in-line with SimCube.

SPECIFICATIONS:
Physical Dimensions   
Size : 3"x 3"x 3.5" (7.6cm X 7.6cm  X 8.9cm)

Weight : 2.5 Lbs

Power : External A/C Adaptor

NIBP Connection : Quick Disconnect 

ECG/Resp Connection : 10 ECG snaps

Manometer Range : 0 – 480 mmHg 
Precision : 0.5 mmHg 
Accuracy : +/-1% of reading

User Interface : Single Button Operation 
   
Operating Modes  :
Manometer 
Adult NIBP
Neo NIBP
Hypertensive NIBP
ECG Asystole
ECG Resp OFF
ECG Pace ON

NIBP Adult Simulation   
Simulated Pressure : 120/80 (100) mmHg 
Simulated Heart Rate : 70 bpm 
Simulated Pulse Volume : 1 ml

NIBP Neonatal Simulation   
Simulated Pressure : 70/40 (55) mmHg 
Simulated Heart Rate : 95 bpm 
Simulated Pulse Volume : 0.5 ml

NIBP Hypertensive Simulation   
Simulated Pressure : 190/120 (150) mmHg 
Simulated Heart Rate : 70 bpm 
Simulated Pulse Volume : 1 ml

ECG Simulation   
Isolated : Yes 
R Wave Size : 1mV (lead I) +/- 5% 
R Wave Width : 35 ms 
Wave Shape : R wave only 
Connection : 10 Snaps 
Rate : 70 bpm (94 bpm for neonatal)

Pacer Simulation :
Isolated : Yes 

Pacer Size : 3 mV 
Pacer Width : 1.2 ms

Respiration Simulation  
Isolated : Yes 
Wave Shape : Square Wave 
Size : 4 Ohm 
Rate : 35 bpm (47 bpm for neonatal)

Environmental   
Voltage Range : 100-240 VAC, 50-60 Hz