Pronk Technologies FT-1KIT Flotrax Infusion Pump Analyzer

Pronk Technologies FT-1KIT Flotrax Infusion Pump Analyzer
Pronk Technologies FT-1KIT Flotrax Infusion Pump Analyzer Pronk Technologies FT-1KIT Flotrax Infusion Pump Analyzer Pronk Technologies FT-1KIT Flotrax Infusion Pump Analyzer Pronk Technologies FT-1KIT Flotrax Infusion Pump Analyzer
Call for Price
Brand: Pronk Technologies
Product Code: FT-1KIT
Availability: In Stock
SALES LEAD TIME: 14 DAYS (Built to Order) (SALES USA Only at this time)
WARRANTY: 4 YEARS OUTRIGHT
ACCESSORIES: Soft Carrying Case, Cleaning solution, Cleaning Brush, 2 AA Batteries, Priming Syringe, Temperature Proble Extension, USB Cable 3 way stopcock, Drain Tubing, Operator Manaul, Calibration Certificate
OPTIONAL ACCESSORIES: IV Pole Clamp
 
INTRODUCTION:
The first palm-size Infusion Pump Analyzer, FlowTrax® has numerous features in a very small affordable and portable package — while rivaling the larger, more expensive competitive products with impressive accuracy and speed. IV pumps can now be tested for Flow Rate and Volume in 3 minutes or less within a 1% accuracy using FlowTrax® patent-pending fluid tracking technology. FlowTrax® includes all the features of more expensive larger IV Pump Analyzers, a full featured pressure/temperature meter and a digital stopwatch. FlowTrax® comes standard with a 4-year parts and labor warranty.
UNIQUE FLUID DETECTION SYSTEM
FlowTrax® was designed with a patent-pending dynamic fluid detection system which enables FlowTrax® to track fluid movement in real time. This measurement system provides 1200 measurement points in a 1ml fluid column, thus providing superior accuracy in fluid volume and rate calculations over traditional analyzers. 
 
SPECIFICATIONS
Accuracy: 1% of reaading +/- 1 LSD for 1ml/hr to 999ml/hr, 2% +/- 1 LSD below 1ml/hr. 0.85% of reading for 500ml/hr, 12ml volume under steady flow conditions.
Range: -28 degree to 115.5 degrees Celsius
Resolution: Flow Rates: 0.01ml; Volume: 0.001ml
Occlusion Pressure Testing: Provides real time and peak pressure measurements.
Digital Timer / Stopwatch:  Automatically starts and stops with fluid measurements. No need to stop analyzer when test is complete.
 
Integral Temperature Meter Functions:
Range: -28.9 to 115.5 degrees Celsius
Units of Measure: Fahrenheit, Celsius
 
Pressure Performance
 
Range: -12.5 to 75 PSI
Accuracy: 0.5% of reading +/- 1 LSD
Resolution: .01 PSI
Units of Measure: mmHg, PSI (G), inH20, cmH2O, kPa, atm, inHg, BAR (A), BAR (G)
 
Flow and Volume Performance
 
Range: 0.5 to 999 ml/hr
Accuracy: 1% of reading +/- 1 LSD for 1ml/hr to 999ml/hr, 2% +/- 1 LSD below 1ml/hr. 0.85% of reading at 500ml/hr, 12ml Volume under steady flow conditions
 
Required for Accuracy:
 
Flow Rate:                             Required Volume Time:
.5 to 10ml/hr                          0.5ml
10.1 to 199ml/hr                   3 minutes (see 3 minute quick testing quick ref. chart)
200 to 999ml/hr                    10ml
 
Precision
Volume: ≤ 0.01ml Precision of FlowTrax is equivalent to 0.01 g based on the conversion of volume (ml) to weight (g) by calculation using standard temperature and density value of water. Precision was established using a rate of 500ml/hr and a “dispensed” volume of 10ml. 
 
Temperature Performance
Range: -28.9 to +115.5 degrees C
Accuracy: 0.1% +/- 0.2 degrees C
Resolution: 0.1 degrees C
 
Stop Watch
Range: 0 to 12 Hours
Accuracy: 0.5%
 
General
Power: Powered by two AA (Alkaline or Lithium) cells or 6 VDC Input
Battery life: Alkaline – Minimum 4 hours, Lithium – Minimum 12 hours
Size (HxWxD): 5.5 in x 3.25 in x 2 in
Operating Temperature: 10 to 35 degrees C (50 to 95 degrees F)
 
 * Environmental condition requirements for Flow Rate and Volume accuracy specifications- Distilled water as test solution, output bottle height of 0 to 18” above FT output, FT horizontal for rates below 21ml/hr, vertical for rates at or above 21ml/hr.